【DIRECTIONS FOR USE】
Operation manual thoroughly before performing the test.
- 1.Before the experiment, the refrigerated reagents and test samples should be balanced to room temperature; the memory is inserted into the instrument,
the instrument automatically recognizes, and the batch information is confirmed to be consistent with the batch on the kit;
- 2.Remove the test card from the original packaging reagent bag, place it on a clean platform, place it horizontally and mark it. It should be used as soon as possible within 1 hour.
- 3.Unseal the buffer and transfer 10μL of the sample into the buffer (See illustration 1) ,insert the head of the tube and mix well (See illustration 2).
- 4.Transfer 6 drops of the buffer (approx.90μL) into the sample well (See illustration 3).
- 5.Wait and read the result with reader (See illustration 4) at 10 minutes. Do not interpret the result after 15 minutes.